Sources for quality control and calibration of nuclear medicine equipment, as well as for defining anatomical positions and organ sizes during SPECT studies. For each type of “accessory” source, different nuclides, activity levels, and geometries are available.
Standard sources for dose calibrators provide a safe and convenient method for performing quality controls and calibration of dose calibrators. The dose calibrator vial is produced by dispersing the active element in 20 ml of epoxy resin, with a density of approximately 1.0 g/cm³. Each standard is supplied in a 27 ml polyethylene vial, and the calibration is expressed in terms of the activity contained in the aqueous solution (NIST traceable calibration certificate).

The Pen Point Marker is used to indicate a point or region of interest during SPECT studies. This source contains Co-57 embedded in a resin matrix located at the tip of a 13.7 cm anodized aluminum rod. The pen-shaped rod is screwed into a brass shielded holder. The activity of the supplied source is within ±15% of the nominal value.
Rulers and radioactive markers are used to define anatomical positions and/or organ dimensions during SPECT studies.
The activity of these sources is within ±15% of the nominal value.

Spot Markers are used for patient positioning during SPECT studies. The active area, with a 3.45 mm diameter, is marked with purple epoxy resin and located at the center of a transparent acrylic disc (25.4 mm × 6.35 mm). The activity of the supplied source is within ±15% of the nominal value.
The Flexible Ruler is a plastic tube with a diameter of 4.8 mm, containing 45 alternating 10 mm sections of Co-57 and inactive lead. The total length of the ruler is 48 cm, with an active length of 45 cm. 
Co-57, uniformly distributed in an epoxy matrix, is contained within a flexible plastic tube with an outer diameter of 2.4 mm and an inner diameter of 0.79 mm. The total length is 50 cm, and the active portion of the ruler measures 48.7 cm.
The Rigid Ruler is a plastic tube with a 5 mm diameter, containing 15 alternating sections of Co-57 and inactive lead.
The total length of the ruler is 17 cm, with an active length of 15 cm. 
Each source consists of a polypropylene vial containing 0.75 ml of active epoxy resin, while the remaining volume is filled with inactive resin. Each set includes the following isotopes: Ba-133, Cs-137, Co-57, Co-60, Cd-109, Mn-54, and Na-22.
The vials are available in two sizes:
T1) 55 mm × 12 mm
T2) 75 mm × 12 mm
Other nuclides and vial sizes are available upon request. Customer-supplied vials can also be used. All Gamma Tube Standard sources are calibrated (NIST Traceable Calibration Certificate).
Each source consists of a lucite rod, available in two sizes:
R2) 127 mm × 12.7 mm
R3) 74.9 mm × 12.7 mm
The active material is sealed in a 4.75 mm × 4.75 mm capsule located at one end of the rod. Each set includes Ba-133, Cs-137, Co-57, Co-60, Cd-109, Mn-54, and Na-22. All Rod Sources are calibrated (NIST Traceable Calibration Certificate).